z.one ultra and z.one ultra sp OPERATOR MANUAL 7-41
Frequency
SmartCart
SmartCart sp
Scan Engine
Press Frequency button to cycle through
the transmit frequency choices.
Highlight Frequency. Press Left / Right
function on Menu Control to increase or
decrease transmit frequency for Color/
Power Doppler. Current frequency is
displayed in Image Information area.
Persistence Press the Menu button. Highlight Persistence with Trackball. Move Trackball left/right
(or press left/right Set buttons) to select preferred level of M-Mode line averaging.
Edge Press Menu button. Highlight Edge with Trackball. Move Trackball left/right (or press
left/right Set buttons) to desired image edge enhancement option.
A Output Press Menu button. Highlight A Output with Trackball. Move Trackball Left to decrease
transmit power output level. Move Trackball Right to increase transmit power output
level. Current transmit power level is displayed (as a percentage; 25%, 50%, etc.) in
Image Information area.
NOTE: Minimum is a power level of 25%. Default transmit power output level is 100%.
Table 7-3. Color Doppler and Power Doppler Controls (Continued)
CD/PD Control SmartCart/SmartCart sp Scan Engine
Indicates that this control is also available in retrospective processing