6
z.one ultra and z.one ultra sp OPERATOR MANUAL 6-17
NOTE: Consult the manufacturer’s Instructions For Use before using a TEE
cleaner or disinfectant. For the latest list of TEE transducer disinfectants approved
by
ZONARE, go to www.zonare.com/products/accessories.
Cleaners (Wipe
Method)
Disinfectants
(Immersion
Method)
Leakage
Safety Testing
The leakage safety test is performed to evaluate the integrity of the transducer
insulation that may become damaged from cuts or bites in the flexible shaft. Cuts
or bites may cause serious electrical leakage that can be hazardous to the
patient. Fluid can enter through the cuts and may cause mechanical and/or
electrical problems.
NOTE: ZONARE recommends doing a leakage test after each patient exam to
ensure patient safety.
Cidex® is a good electrically conductive medium that works well for testing the
TEE transducer in the leakage tester. If Cidex or another disinfecting agent is
used in the tester setup a Cidex-compatible soak tray should be used. BC
Biomedical, the manufacturer of the leakage tester, has a Cidex-compatible soak
tray available for purchase (BC Biomedical part number: BC20-42200). If Cidex is
not used for the testing, a bath of 0.7% saline solution is also a suitable
conductive solution for testing electrical leakage.
Cidezyme®/Enzol® Prolystica® 2x conc. Enzymatic Presoak & Cleaner
Empower® T-Spray™
Hycolin® T-Spray II™
MetriZyme® Wexcide 128
Anioxyde 1000 Omnicide 14 NS®
Cidex ADS® Omnicide 28®
Cidex Plus® Perasafe®
Cidex OPA® Steranios 2%®, 2% NGV, 2% ECS®
Gigasept AF® TD-100™ & TD-5™
Gigasept FF® Totacide 28
Korsolex® extra Tristel One Day Concentrate (T1DC)
Metricide® Virkon®
Metricide 28® Wavicide-01®
Metricide Plus 30® Wexcide 128
Metricide OPA Plus®