11-10 z.one ultra and z.one ultra sp OPERATOR MANUAL
IQ Scan Data
The IQ Scan feature allows you to store raw source (prescan converted/
processed) ultrasound data to the internal storage and make retrospective
adjustments to 2D, Color, and spectral Doppler. After these adjustments are
made, you can restore the newly adjusted image.
However, due to the raw ultrasound data process involved in IQ Scan image
capture, storage space on the internal archive will fill up more quickly and image
store times may be noticeably longer.
IQ Scan data is not intended for permanent patient exam of record image
archival. The normal exam of record comprises standard images, archived in
DICOM format, that are maintained and supported for viewing across all software
releases.
NOTE: Archived IQ Scan data images cannot be guaranteed to be compatible for
importing back into the System due to potential compatibility conflicts of new
software releases.
To Store IQ Scan Data
Go to Image Store/Print Button screen (Figure 11-4) and check box for IQ Scan
Data and/or 3D Volume Data
To Store 2D and Color Doppler Clips - IQ Scan
Still Images
1. Freeze the 2D or Color Doppler image.
2. Press Store button associated with IQ Scan.
Clips
1. Freeze the 2D or Color Doppler image.
2. Press Menu button and activate Cine.
3. While in Cine playback, press the Store button. The length of the stored clip
was defined when you configured Clip Storage Duration using the Protocol
button (see
“Clip/Image Store (Protocol Button - SmartCart only)” on
page 7-12).
NOTE: If desired, trim the Cine before storing the clip.
4. Go into Current Exam, press Menu, and select IQ Scan. The IQ Data will
load.
5. Press Freeze to run clip. Depending on the mode you are in, you can adjust
various 2D, Color Doppler parameters while clip is playing back.
SmartCart/SmartCart sp
Has one Store button and one Print button for
storing IQ Scan Data
Scan Engine
Has one Store button and one Print button for
storing IQ Scan Data