Quatech Audio/Video Capture USB 2.0 Dokumentacja Strona 369

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 368
15
z.one ultra and z.one ultra sp OPERATOR MANUAL 15-3
Image too soft or hazy
Increase Edge setting
Decrease Persistence setting
Image too much contrast
Increase Dynamic Range setting
Decrease Map setting
Use Compounding mode
Image grainy
Decrease Edge setting
Increase Frequency
Increase Persistence setting
Use Compounding mode
Insufficient resolution
Increase Frequency
Zoom to area of interest
Use Optimize sound speed correction (ZST)
Use a higher frequency transducer for the application
Insufficient penetration
Decrease Frequency
Use a lower frequency transducer for the application
Lost part of image
Make sure gel covers complete face of transducer
Make sure transducer is in complete contact with patient
If a transducer cover is used, remove air bubbles between cover and
transducer face
No (or poor quality) B-Mode Image
Make sure the brightness and/or contrast settings for the LCD Display have
not been altered from proper settings. For information about LCD Display
settings, see
page 2-6.
NOTE: To ensure the quality of the grayscale resolution of the LCD Display, use
the grayscale test pattern (via the Diagnostic menu) available on the
SmartCart/
SmartCart sp
. Ensure that all shades can be visualized, making small
adjustments to brightness and contrast from factory defaults, as required, to see
all intensities.
Adjust the B-Mode Gain control.
In the onscreen menu, adjust the A Output (acoustic power) level.