9
9-24 z.one ultra and z.one ultra sp OPERATOR MANUAL
NOTE: Acceleration is always calculated in the direction of positive time. The
velocity caliper positioned at the earliest strip time is always the startpoint (and the
other caliper is the endpoint) for calculating the delta velocity (DeltaV), delta time
(AT), and acceleration (Accel) results.
NOTE: Pressing the Set button again toggles the active caliper selection between
the first and second velocity calipers, allowing positional corrections to be made to
both caliper instances within the strip image.
8. Press the Measure button to remove the Accel tools and exit Measure mode.
Systolic/
Diastolic Ratio
The PW Systolic/Diastolic (S/D) Ratio is defined as abs (Max/Min), where Max
and Min are determined by comparing the absolute velocities obtained by two
concurrent PW Velocity calipers positioned within the same PW Doppler strip.
Only a single instance of this tool can exist at any one time.
•The DDB displays both Velocity results and their associated S/D calculation
result and continuously updates the results when the user positions either
caliper within the PW Doppler strip image.
In the DDB, Max velocity is identified as PSV (Peak Systolic Velocity) and Min
velocity is identified as EDV (End Diastolic Velocity).
To Measure Systolic/Diastolic (S/D) Ratio
1. Freeze the display with a PW Doppler strip image.
2. Press the Measure button.
3. Highlight and select S/D.
4. The PSV velocity caliper will be displayed.
5. Use the Trackball to position the caliper to the desired location within the PW
Doppler strip display. At this time:
6. Press the Set button to fix the PSV velocity caliper and activate the EDV
velocity caliper.
NOTE: The S/D result is calculated and displayed once both velocity calipers have
been activated.
7. Use the Trackball to position the caliper to the desired location within the PW
Doppler strip display.
NOTE: If the absolute value of the EDV velocity result is greater than the current
PSV result, the EDV and PSV result displays are swapped and the caliper becomes
a PSV caliper. The converse is also true when the absolute value of the PSV
velocity result is less than the current EDV result.
SmartCart/SmartCart sp
Only displays the PSV velocity result (in the DDB)
and the System updates the result as the user
positions the caliper within the strip image
SmartCart/SmartCart sp
Only updates the EDV velocity result and the S/D
calculation result as the user positions the caliper
within the strip image