z.one ultra and z.one ultra sp OPERATOR MANUAL 7-57
Figure 7-24. 3D Setup Mode Status Field
5. To adjust acquisition quality (QA in the status field), press Quality softkey to
cycle through options: HIGH, BALANCED, and FAST.
NOTE: The default is HIGH, which results in highest resolution but slowest
acquisition time. Choose BALANCE or FAST if faster acquisition times are desired.
6. To adjust elevation sweep angle (E in status field), press Angle softkey; each
press changes sweep angle, which is distance across sweep path. The
sweep angle choices are 75°; 60°; 45°; 30°; 20°. Default is 60°.
7. Place ROI in position using Trackball and Set button. Position ROI so that the
cut line (yellow) is in front of fetal face with as much amniotic fluid in between
face and cut line.
Figure 7-25. 3D ROI with Cut Line
8. Press Freeze button. This initiates image acquisition. A progress bar indicates
status of acquisition.
NOTE: You can also initiate acquisition by pressing far-left softkey labeled Start/
Stop.
NOTE: To abort acquisition, press Freeze or Start/Stop key to return to 3D Setup
mode.
Quality is HIGH
Elevation Sweep
Angle is 60
º
Cut line