9
9-40 z.one ultra and z.one ultra sp OPERATOR MANUAL
6. A second caliper tool becomes active for the next Hip Angle measurement
component.
Bony Roofline
1. Use the Trackball to position the green caliper tool at the beginning of bony
roofline.
2. Press Set to lock the position of the caliper tool. A second caliper tool
becomes active.
3. Use the Trackball to position the second caliper tool along the iliac baseline. A
connecting line is drawn between the two calipers and is sized as the first
caliper is positioned and the intersection with the iliac baseline is set by the
second caliper position.
4. To fine-tune the bony roofline, use the Set button to toggle the active caliper
selection.
5. To lock the position of the bony roofline:
6. A second caliper tool becomes active for the next Hip Angle measurement
component.
Cartilage Roofline
1. Use the Trackball to position the green caliper tool at the beginning of
cartilage roofline.
2. Press Set to lock the position of the caliper tool. A second caliper tool
becomes active.
3. Use the Trackball to position the second caliper tool along the iliac base line.
A connecting line is drawn between the two calipers and is sized as the first
caliper is positioned and the intersection with the iliac baseline is set by the
second caliper position.
4. To fine-tune the cartilage roofline, use the Set button to toggle the active
caliper selection.
5. To lock the position of the cartilage roofline and complete the measurement:
6. The Hip Angle measurement results are copied to the Worksheet. The data
display in the Measurements window inverts to black on white.